Capsule endoscopy in patients with cardiac pacemakers, implantable cardioverter defibrillators and left heart assist devices (EN)

Höltgen, Reinhard
Bandorski, Dirk
Keuchel, Martin
Stunder, Dominik

According to the recommendations of the US Food and Drug Administration and manufacturers, capsule endoscopy should not be used in patients carrying implanted cardiac devices. For this review we considered studies indexed (until 30.06.2013) in Medline [keywords: capsule endoscopy, small bowel endoscopy, cardiac pacemaker, implantable cardioverter defibrillator, interference, left heart assist device], technical information from Given Imaging and one own publication (not listed in Medline). Several in vitro and in vivo studies included patients with implanted cardiac devices who underwent capsule endoscopy. No clinically relevant interference was noticed. Initial reports on interference with a simulating device were not reproduced. Furthermore technical data of PillCam (Given Imaging) demonstrate that the maximum transmission power is below the permitted limits for cardiac devices. Hence, impairment of cardiac pacemaker, defibrillator or left ventricular heart assist device function by capsule endoscopy is not expected. However, wireless telemetry can cause dysfunction of capsule endoscopy recording. Application of capsule endoscopy is feasible and safe in patients with implanted cardiac devices such as pacemakers, cardioverter defibrillators, and left heart assist devices. Development of new technologies warrants future re-evaluation.Keywords Capsule endoscopy, cardiac pacemakers, implantable cardioverter defibrillator, interference, left heart assist deviceAnn Gastroenterol 2014; 27 (1): 3-8 (EN)

info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion


Αγγλική γλώσσα

2013-12-04


Hellenic Society of Gastroenterology (EN)

1792-7463
1108-7471
Annals of Gastroenterology; Volume 27, No 1 (2014); 3 (EN)




*Η εύρυθμη και αδιάλειπτη λειτουργία των διαδικτυακών διευθύνσεων των συλλογών (ψηφιακό αρχείο, καρτέλα τεκμηρίου στο αποθετήριο) είναι αποκλειστική ευθύνη των αντίστοιχων Φορέων περιεχομένου.