Design of experiments for a precise estimation of the calibration curve of periodate in its determination by flow-injection analysis and chemiluminescence detection.

 
This item is provided by the institution :

Repository :
Repository of UOI Olympias
see the original item page
in the repository's web site and access all digital files if the item*
share




1995 (EN)

Design of experiments for a precise estimation of the calibration curve of periodate in its determination by flow-injection analysis and chemiluminescence detection. (EN)

Papamichael, E. M. (EN)

Πανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείας (EL)
Papamichael, E. M. (EN)

A design of experiments method based on the criterion of D-optimality is presented. It is applied to the estimation procedure of the parameters of a periodate calibration curve described by a model equation of the form y = ax(2)/(bx + c root x + 1) and giving the maximum possible precision using three experimental measurements. The use of four or five experimental measurements for producing the calibration curve was not found more advantageous than that of three. Simulated data were generated, for the three optimal points, and a large number of parameter estimates were obtained under constant and/or relative errors (in y) showing a robust behaviour. The %-bias and the %-excess variance based on the simulated data were found to be non-significant. (EN)

experimental design (EN)


Chemometrics and Intelligent Laboratory Systems (EN)

English

1995


Elsevier (EN)




*Institutions are responsible for keeping their URLs functional (digital file, item page in repository site)