Sol-Gel Glass Immunosorbent-Based Determination of s-Triazines in Water and Soil Samples Using Gas Chromatography with Nitrogen Phosphorous Detection System (EN)

Stalikas, C. D. (EN)

Πανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείας (EL)
Stalikas, C. D. (EN)

A rapid and efficient method for the selective extraction of s-triazine herbicides in environmental samples was developed using an immunosorbent of monoclonal antiatrazine antibodies, which were encapsulated in a sol-gel glass matrix. The cross-reactivity of the antibody for analytes structurally related with atrazine enabled the simultaneous extraction of several s-triazine herbicides (atrazine, propazine, terbuthylazine, cyanazine, desethyl atrazine). After trace enrichment on the immunoextraction column, the s-triazines were desorbed by means of an acidic buffer (pH 2.5) and further extracted with ethyl acetate before being injected into the GC. Compared to liquid-liquid extraction and solid-phase extraction with a hydrophobic SDB-L support, the GC-NPD chromatograms obtained after immunoaffinity enrichment of surface water (river water) samples or soil extracts and analysis were free from matrix interferences. Nonspecific adsorption of humic acids was not observed. The method allows for the determination of the herbicides in linear ranges up to 1.5 microg/L with correlation coefficients higherthan 0.99 and relative standard deviations between 4% and 7% (n = 5). The LODs for 50-mL water samples were in the range 0.02 microg/L (atrazine, propazine) to 0.1 microg/L (desethyl atrazine) (S/N = 3). In addition to its high selectivity, the immunosorbent proved to be reusable for a significant number of preconcentration runs. However, the composition of samples may influence the lifetime of the column. (EN)


Environ. Sci. Technol. (EN)

Αγγλική γλώσσα

2002


American Chemical Society (EN)




*Η εύρυθμη και αδιάλειπτη λειτουργία των διαδικτυακών διευθύνσεων των συλλογών (ψηφιακό αρχείο, καρτέλα τεκμηρίου στο αποθετήριο) είναι αποκλειστική ευθύνη των αντίστοιχων Φορέων περιεχομένου.