A new γ-spectroscopy station at the University of Athens

see the original item page
in the repository's web site and access all digital files if the item*



A new γ-spectroscopy station at the University of Athens (EN)

Kouvaris, E.
Mertzimekis, T. J.
Lagaki, V.

info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion

2019-03-08


A new γ-spectroscopy station at the University of Athens has been recently deployed. The station is built around a 25% HPGe detector and aims at supporting the environmental radioactivity studies program of the NuSTRAP group at UoA. The detector needs detailed characterization of its efficiency and calibration over a rather wide range of energies. Besides standard calibration point sources (60Co, 137Cs, 152Eu and 226Ra), detector simulations using the Monte Carlo N-particle transport code (MCNP) were performed. Results of the MCNP calculations are shown and compared with real spectra showing satisfactory agreement (EN)


efficiency calibration (EN)
γ-ray spectroscopy (EN)
MCNP (EN)
HPGe detector (EN)

Annual Symposium of the Hellenic Nuclear Physics Society

English

Hellenic Nuclear Physics Society (HNPS) (EN)


2654-0088
2654-007X
Annual Symposium of the Hellenic Nuclear Physics Society; Τόμ. 22 (2014): HNPS2014; 126-128 (EL)
HNPS Advances in Nuclear Physics; Vol. 22 (2014): HNPS2014; 126-128 (EN)

Πνευματική ιδιοκτησία (c) 2019 V. Lagaki, E. Kouvaris, T. J. Mertzimekis (EL)