Electron fluence determination per Monitor Unit of a linear accelerator used for radiation therapy

see the original item page
in the repository's web site and access all digital files if the item*



Electron fluence determination per Monitor Unit of a linear accelerator used for radiation therapy (EN)

Makridou, A.
Zamani, M.
Tziaka, C.
Fragopoulou, M.
Stoulos, S.

info:eu-repo/semantics/article
info:eu-repo/semantics/publishedVersion

2019-11-23


About 20% of patients in Greece undergoing radiotherapy were irradiated by electron beams with energies ranged between 4 to 20 MeV. The correct determination of electron beam, leads to the proper treatment planning. In this study electron fluence at the isocenter was determined for a linear accelerator ELECTA SL 20 for nominal electron energy of 20 MeV. Two methods have been applied using (a) an ionization chamber in the frame of monthly quality assurance and activation detector (238U). The experimental estimated electron flux, which was in good agreement with the applied calculations using the data of irradiation procedure, ranged between 3.5 up to 4.3 x 108 electrons per cm 2 per sec or 6.1 ±0.7 x 10^7 e.cm-2 per MU. (EN)


Annual Symposium of the Hellenic Nuclear Physics Society

English

Hellenic Nuclear Physics Society (HNPS) (EN)


2654-0088
2654-007X
Annual Symposium of the Hellenic Nuclear Physics Society; Τόμ. 18 (2010): HNPS2010; 204-207 (EL)
HNPS Advances in Nuclear Physics; Vol. 18 (2010): HNPS2010; 204-207 (EN)

Πνευματική ιδιοκτησία (c) 2019 A. Makridou, C. Tziaka, M. Fragopoulou, S. Stoulos, M. Zamani (EL)




*Institutions are responsible for keeping their URLs functional (digital file, item page in repository site)