Functionalization of niobium electrodes for the construction of impedimetric biosensors

 
This item is provided by the institution :

Repository :
Repository of UOI Olympias
see the original item page
in the repository's web site and access all digital files if the item*
share




2008 (EN)

Functionalization of niobium electrodes for the construction of impedimetric biosensors (EN)

Helali, S. (EN)

Πανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείας (EL)
Helali, S. (EN)

This paper describes the development of an impedimetric immunosensor, based on niobium/niobium oxide (Nb[NbOxHy) electrodes, for the detection of atrazine. Niobium oxide was anodically formed onto niobium electrodes at 25 V in 1 M H2SO4. Hydrous oxide layers were then silanized with APTES, and using glutaraldehyde as a cross linker, Fab fragment k47 antibody was covalently immobilized onto the surface of the electrodes. Electrochemical impedance spectroscopy (EIS) was used to characterize the building-up of the immunosensors as well as the binding of atrazine to its specific antibody. In presence of ferricyanide redox species and under a cathodic polarization voltage (-1.2 V versus SCE), the relationship between the concentration of atrazine and the change of the electron transfer resistance value was studied. (c) 2007 Elsevier B.V. All rights reserved. (EN)

atrazine (EN)


Materials Science & Engineering C-Biomimetic and Supramolecular Systems (EN)

English

2008


Elsevier (EN)




*Institutions are responsible for keeping their URLs functional (digital file, item page in repository site)