An enzymic method for the determination of bilirubin using an oxygen electrode

 
This item is provided by the institution :
University of Ioannina
Repository :
Repository of UOI Olympias
see the original item page
in the repository's web site and access all digital files if the item*
share



2000 (EN)
An enzymic method for the determination of bilirubin using an oxygen electrode (EN)

Klemm, J. (EN)

Πανεπιστήμιο Ιωαννίνων. Σχολή Θετικών Επιστημών. Τμήμα Χημείας (EL)
Klemm, J. (EN)

An enzymic method for the determination of bilirubin is described, based on the indirect electrochemical monitoring of bilirubin via its reaction with oxygen to biliverdin in the presence of bilirubin oxidase. Amperometric signals due to oxygen depletion were taken by means of an oxygen electrode incorporating an immobilized bilirubin oxidase membrane. The calibration plot shows linearity up to 2.0 x 10(-4) M. This concentration range corresponds to elevated bilirubin concentrations in adults and in newborn infants. The RSD of the method is 3.1 % (1.0 x 10(-4) M bilirubin, n = 6). The method is interference free, fast ( < 3min) and easy to perform. The membrane retains almost 80 % of its initial activity after 30 successive measurements. (EN)

bilirubin (EN)

Πανεπιστήμιο Ιωαννίνων (EL)
University of Ioannina (EN)

Electroanalysis (EN)

English

2000

<Go to ISI>://000085963100009

WILEY-VCH Verlag (EN)