δείτε την πρωτότυπη σελίδα τεκμηρίου στον ιστότοπο του αποθετηρίου του φορέα για περισσότερες πληροφορίες και για να δείτε όλα τα ψηφιακά αρχεία του τεκμηρίου*
From disk devices to transcatheter patches: The evolution of wireless heart defect occlusion
Sideris, E.B.
Sideris, C.E.
Savvas, T.
Moulopoulos, S.D.
Since 1988, we have been using double-disk devices with several device generations and improving results. Our current disk device, "The ButtonSeal Centering on Demand Device," is a multipurpose device made of stainless steel and polyurethane. It has shown high occlusion rates and no significant complications in atrial septal defects (ASDs) up to 30 mm in diameter. It can be used according to the ASD anatomy both in large, single defects (centering) and in multiple fenestrations (noncentering). The ButtonSeal shares the same limitations as other disk devices, namely, the possibility, of wire-related problems and the need for significant septal rim. The development of wireless balloon-delivered devices and patches addresses these problems and limitations. Both detachable balloon devices and patches were used experimentally in animals and in feasibility studies in humans. The results showed good occlusion rates and lack of wire-related complications. Both methods required a minimal rim and occluded more defects than the disk devices. The transcatheter patch method was safer and had only one disadvantage, which was the need for 48-hour balloon support.
(EN)
*Η εύρυθμη και αδιάλειπτη λειτουργία των διαδικτυακών διευθύνσεων των συλλογών (ψηφιακό αρχείο, καρτέλα τεκμηρίου στο αποθετήριο) είναι αποκλειστική ευθύνη των αντίστοιχων Φορέων περιεχομένου.